Prepared | K. Duffek, J. Hoscheidt | 12-May-21 |
Reviewed
Approved
Issue | Date | Changes | Responsible |
a | 12-May-21 | Initial Release | K. Duffek, J.Hoscheidt |
VATTEL VME computer controlling the Telescope
Neslab Primary Mirror cooling system
Purpose
The purpose of this document is to provide a detailed startup/shutdown procedure for remote operation of the Vatican 1.8m Telescope.
Reference Documents
[RD1] Vatican 1.8m Telescope Operators Manual.
[RD2] Lightning shutdown procedure.
General
Verify all doors are locked.
Close all blinds, shut off unneeded lights, turn off TV.
Computer Room
Power up VATT computer rack beginning with VATT Data computer.
Wait 20 minutes for VATT Data to boot up.
Then boot all other rack computers in no particular order.
frame2
Dome
Check that filters have been installed in the filter wheel and their positions recorded for the observer.
Top off the science camera with LN2.
Remove Dome Short.
Plug in Dome Shutter power into 120ac outlet.
Plug in finder scope camera power located on top of grey box on right fork.
Turn 4Kccd camera power ON.
Turn ON 4Kccd camera shutter box.
Turn ON guide camera power at the guide camera controller.
Remove finder scope cover.
Plug in stow pin power supply located on left fork. Switch ON.
Remove stow pins pressing “unlock” on momentary switch until display reads 100 approximately.
Visually verify the pins are out. Switch OFF, then unplug.
Make sure Telescope is free to move and is clear of any obstacles.
Pier Stairwell Second floor
VATTEL Room
Plug in GPS antenna, power unit up and wait for GPS lock before proceeding to next steps.
Power up VATTEL.
Power up VATT Guider.
Power up VATT Secondary.
Flip ON all drive amplifier power switches (except counter weights) and observe current meters are OFF except for the Dome amp.
Basement
Power up building exhaust fans and pier fans in the oil pump room.
Read oil temperature; it should be 61 degrees Fahrenheit.
Power up oil pump following the front panel steps.
Wait for chiller to come up and allow a few minutes to mix oil. Watch for jump in temperature reading.
Watch gauge pressure rise to operating range. Make sure oil pressure reads 1,180 psi with no apparent oil leaks.
Verify green lights on panel by door.
Dome
Control Room
Power ON the Dome, All sky, and Finder monitors as well as Dome Intercom (Note dome 1 button flakey; dome 2 cannot hear).
Make a Skype call to the Observer letting them know the Telescope is ready for observing.
Power up the Dome Fan after the Dome Slit and Mirror Cover has been opened. The controller is on the south wall to the right of the back door.
Give observers temp, humidity, wind speed and wind direction from weather station when open (note to talk to Tom Folkers about getting info from SMT)
Control Room
Turn Dome Fan OFF.
Have observer stow scope, close mirror covers, close dome, home dome and kill amplifiers.
Give observers temp, humidity, and wind speed and wind direction from weather station upon closing.
Turn OFF Guider, Finder and Sky monitors.
VATTEL Room
Turn OFF all amplifier power switches: Azimuth Servo, Derotator, Elevation Servo, Dome
Verify that all current meters are OFF.
Pier Stairwell, Second Floor
Shut gate valve on right side of Neslab until almost closed then immediately switch Neslab power OFF and close gate valve all the way.
Verify red lines on Neslab pneumatic valve change from parallel to perpendicular.
Basement
Dome
Verify dome closed!
Plug in stow pin power supply. Switch ON.
Engage stow pins pressing “lock” on momentary switch until display reads 000 approximately.
Visually verify the pins are in. Switch OFF, then unplug.
Turn Guider camera OFF.
Close finder scope cover.
Turn OFF 4Kccd camera shutter power supply (leave on if observations continue).
Turn OFF 4Kccd camera power supply (leave on if observations continue).
Unplug finder camera power located on top of grey box on right fork.
Unplug Dome Shutter power.
Install Dome Short.
Check dome fiduciary mark on wall against position of bolt on underside of dome ring and note the error; if error greater than 5º, have Gary reset Dome 0.
Fill Dewar (if needed) and record fill time on clipboard.
Turn OFF lights in dome.
Control Room
Turn OFF Dome Flood light.
Turn OFF Dome monitor.
Turn intercom OFF.
End Skype call with observer.
Lock all doors that may have been unlocked during the night before going to bed.
–oOo–
Doc_info_start
Title: VATT Telescope startup/Shutdown
Document Type: Checklist
Source: Vatican Observatory
Issued by: K. Duffek
Date_of_Issue: 12-May-21
Revised by:
Date_of_Revision:
Checked by:
Date_of_Check:
Accepted by:
Date_of_Acceptance:
Released by:
Date_of_Release:
File Type: MS Word
Local Name: VATT Telescope startup/Shutdown
Category: ##?
Sub-Category: ##?
Assembly: Fill in Assembly name here
Sub-Assembly:
Part Name:
VATT Archive: ##?
Revision: a
Doc_info_end