====VATT Spectrograph==== * This optical region spectrograph has AR coated refractive optics, with a beam size at its camera of 125 mm, giving … * slit length: 30 arcsec * wavelength range: 360 - 950 nm * spectral coverage: 100 nm at 0.1 nm resolution * spectral resolutions: 0.1, 0.2, and 0.4 nm with 1 arcsec wide slit * detector: back illuminated, low noise CCD STA0520A, with very nice cosmetics. STA0520A CCD 2688×512 pixels, 15 microns ITL Serial Number 8228 Detector operating temperature -110C Gain = 1.3 e/DN with setting 10 Gain = 2.6 e/DN with setting 5 For both gains the full well is ADC limited (65k DN), so this is 85,000 or 170,000 electrons. Noise is 3.4 electrons (3.2 in overscan region) * [[VATT:vatt-vattspec_efficiency|efficiency]]: determined by Mark Wagner for telescope plus spectrograph * further details are under the {{ :vatt:vattspec-instructions.pdf|General Instructions}} for use of the VATT Spectrograph. * bandpass [[vatt:filters|filters]] available: S8612 (red blocking); GG-400, GG-475, OG-550 (cut-offs to the blue in nm). ===Documentation=== * {{ :vatt:vattspec-instructions.pdf |General Instructions}} * {{ :vatt:vattspec-guider-notes.pdf |Instructions on Guiding}} ===Maintenance=== * Argon Calibration Lamp was replaced on 4/26/2013. * Note: There is some concern the lamps are burning out faster than they should. ===Chip QE Curve=== {{:vatt:qe_vattspec.png|}}