*make sure the ground wire is connected: * first to the mirror cell and * then to the camera *insert the nozzle into the dewar, and attach it using the bungee cord *opening the valve, start filling the camera's dewar with LN2; if using remotely controllable valve follow [[VATT:Jinvoo LN2 valve|instructions in a separate document]] *fill out LN2/camera log – listen for full dewar *cease flow when dewar is full *disconnect LN2 filler hose from camera dewar, move storage dewar off of telescope platform and south, to dome silo wall) *remove camera ground wire and clip it on silo, near storage dewar