 ====Sunnyside 2m chamber description==== ====Sunnyside 2m chamber description====
-    The vacuum chamber diameter: 84” I.D +  * The vacuum chamber diameter: 84” I.D 
-    The vacuum chamber height: 12’ (13’ including hemispherical bottom). +  The vacuum chamber height: 12’ (13’ including hemispherical bottom). 
-    The chamber has one filament ring 71.75” diameter (center of filaments). +  The chamber has one filament ring 71.75” diameter (center of filaments). 
-    The chamber wall to the center of the filaments is 6.125”. +  The chamber wall to the center of the filaments is 6.125”. 
-    The optimum placement of the surface of the mirror in the chamber is 19” below the chamber lid. +  The optimum placement of the surface of the mirror in the chamber is 19” below the chamber lid. 
-    The top of the filaments to the edge of the mirror is 34” (vertical measurement - 72” diameter mirror). +  The top of the filaments to the edge of the mirror is 34” (vertical measurement - 72” diameter mirror). 
-    The top of the filaments to the chamber lid is 53”. +  The top of the filaments to the chamber lid is 53”. 
-    The mirrors are held from the chamber lid, the coated surface faces down. +  The mirrors are held from the chamber lid, the coated surface faces down. 
-    There is one ring of 36 filaments located near the periphery of the chamber, below the mirror. +  There is one ring of 36 filaments located near the periphery of the chamber, below the mirror. 
-    The filaments are made from 3 strand twisted tungsten wire (individual wire size 0.020”), 6 coils per filament, coiled section of filament 2.5” long, coil diameter 0.44” O.D., overall length of tungsten wire 12.625”. +  The filaments are made from 3 strand twisted tungsten wire (individual wire size 0.020”), 6 coils per filament, coiled section of filament 2.5” long, coil diameter 0.44” O.D., overall length of tungsten wire 12.625”. 
-    Deposition material is aluminum, 99.999% purity, 1mm diameter wire, 1.25” or 1.375” long, folded twice, placed on the bottom of each filament loop, total of 216 pieces of aluminum. +  Deposition material is aluminum, 99.999% purity, 1mm diameter wire, 1.25” or 1.375” long, folded twice, placed on the bottom of each filament loop, total of 216 pieces of aluminum. 
-    The tungsten filaments can be used for two or three shoots before they need to be replaced. +  The tungsten filaments can be used for two or three shoots before they need to be replaced. 
-    Mechanical pump; Kinney KDH 250 rotary piston pump, pumping speed 248 (cfm) +  Mechanical pump; Kinney KDH 250 rotary piston pump, pumping speed 248 (cfm) 
-    Roots Blower pump; Kinney KMBV 1300, pumping speed 1300 (cfm) +  Roots Blower pump; Kinney KMBV 1300, pumping speed 1300 (cfm) 
-    (2) Diffusion pumps; Kinney KDP-32 (32” diameter) pumping speed 31000 l/s (baffled) +  (2) Diffusion pumps; Kinney KDP-32 (32” diameter) pumping speed 31000 l/s (baffled) 
-    Holding pump; Stokes Microvac 148-H9, pumping speed 40 (cfm) +  Holding pump; Stokes Microvac 148-H9, pumping speed 40 (cfm) 
-    Rough vacuum pressure is measured with Granville-Phillips Series 275 controllers with Convectron gauge sensors. +  Rough vacuum pressure is measured with Granville-Phillips Series 275 controllers with Convectron gauge sensors. 
-    High vacuum pressure is measured with Granville-Phillips Series 270 controllers with Bayard-Alpert Series 274 ionization gauges. +  High vacuum pressure is measured with Granville-Phillips Series 270 controllers with Bayard-Alpert Series 274 ionization gauges. 
-    The deposition rate and total thickness are measured by a Maxtek TM-100 R, 6.0 MHz quartz crystal microbalance. +  The deposition rate and total thickness are measured by a Maxtek TM-100 R, 6.0 MHz quartz crystal microbalance. 
 + 
 [[.:Sunnyside Chamber Lid Drawings]] [[.:Sunnyside Chamber Lid Drawings]]
  
public/sunnyside/general_information.1586384239.txt.gz · Last modified: 2020/04/08 15:17 by mel