| - | ======Lightning Shutdown, Short Version: Updated | + | ======Lightning Shutdown, Short Version: Updated |
| + | |||
| + | ===Changes coming out of Downtime 2018=== | ||
| + | - We have an " | ||
| + | - BokCCD2 is now the main Azcam computer for both the B&C and 90Prime instruments. | ||
| + | - There is no longer an UPS under the counter feeding console power. | ||
| + | - The weather computer is now password protected, and the password is posted on the monitor. | ||
| Line 15: | Line 21: | ||
| - Turn off IIS readout and unplug | - Turn off IIS readout and unplug | ||
| - Turn off IIS motor controller and unplug | - Turn off IIS motor controller and unplug | ||
| - | - Turn OFF and unplug ' | + | - Turn off and unplug ' |
| - | *Telescope (for other CASS users) | + | *Telescope (for other CASS users [WIFIS, SPOL, SCC, PISCES]) |
| - Make sure everything is turned off and unplugged from orange outlets on the bottom of the telescope | - Make sure everything is turned off and unplugged from orange outlets on the bottom of the telescope | ||
| - Make sure IIS readout and motor controller are turned off and unplugged. | - Make sure IIS readout and motor controller are turned off and unplugged. | ||
| Line 32: | Line 38: | ||
| * Unplug dome mic if in use next to AED | * Unplug dome mic if in use next to AED | ||
| - Observers Workstation | - Observers Workstation | ||
| - | * Shutdown AzCam program, shutdown Azcam computers (BOKCCD2 & BOKCCD4) from start menu, and shutdown AzCam Xterm | + | * Shutdown AzCam program, shutdown Azcam computers (BOKCCD2 & BOKCCD4 |
| * Shutdown programs on Bart, shutdown Bart from Start Menu, and Bart Xterm | * Shutdown programs on Bart, shutdown Bart from Start Menu, and Bart Xterm | ||
| * Shutdown weather computer | * Shutdown weather computer | ||
| Line 41: | Line 47: | ||
| * #6 Bart (should already be off) | * #6 Bart (should already be off) | ||
| * **#7 BokPOP (Please leave BOKPOP turned ON)** | * **#7 BokPOP (Please leave BOKPOP turned ON)** | ||
| - | * Shut down Bokaux computer | ||
| * Turn off flat field lamp controller if in use | * Turn off flat field lamp controller if in use | ||
| * Turn off printer | * Turn off printer | ||
| * Turn off (3) UPS units in bottom of the rack | * Turn off (3) UPS units in bottom of the rack | ||
| - | * Note: The CONSOLE UPS powers the Operators Console, make sure everything in it is turned off first. | + | * Note: The CONSOLE UPS powers the Operators Console, make sure everything in it is turned off first. |
| Line 91: | Line 96: | ||
| * #6 Bart (should already be off) | * #6 Bart (should already be off) | ||
| * #7 BokPOP **(should stay ON)** | * #7 BokPOP **(should stay ON)** | ||
| - | * Shut down bokaux computer | ||
| * Turn off flat field lamp controller if in use | * Turn off flat field lamp controller if in use | ||
| * Turn off printer | * Turn off printer | ||
| * Turn off (3) UPS units in bottom of the rack | * Turn off (3) UPS units in bottom of the rack | ||
| - | * * Note: The CONSOLE UPS powers the Operators Console, make sure everything in it is turned off first. | + | * * Note: The CONSOLE UPS powers the Operators Console, make sure everything in it is turned off first. |
| *Warm Room Cable Closet | *Warm Room Cable Closet | ||
| Line 125: | Line 129: | ||
| *Electronics Computer Rack Area | *Electronics Computer Rack Area | ||
| - | - Turn on (3) UPS units in bottom of the rack | + | - Turn on (3) UPS units in bottom of the rack. On the new Console UPS: You will need to hit the power button, then use the arrow and OK buttons to choose "On Immediately" |
| - Turn on Bokpop (it should already be ON) | - Turn on Bokpop (it should already be ON) | ||
| - Turn on Bart | - Turn on Bart | ||
| - | - Turn on BokAux | + | - Turn on BokAux |
| - Turn on printer | - Turn on printer | ||
| - **90Prime users** | - **90Prime users** | ||
| Line 139: | Line 144: | ||
| - Turn on Azcam Xterm, log into BokCCD2 if using B&C or 90Prime. | - Turn on Azcam Xterm, log into BokCCD2 if using B&C or 90Prime. | ||
| - Start Bart Xterm, login either as Primefocus Observer if you're using 90Prime, or Bok Observer with any other instrument using appropriate passwords. | - Start Bart Xterm, login either as Primefocus Observer if you're using 90Prime, or Bok Observer with any other instrument using appropriate passwords. | ||
| + | - Start weather computer by pressing button on top of the upper left edge. Enter the password posted on the monitor. | ||
| * At the Operators Console: | * At the Operators Console: | ||
| - Make sure the CONSOLE UPS is plugged in and turned ON (should be done already if the above steps were followed) | - Make sure the CONSOLE UPS is plugged in and turned ON (should be done already if the above steps were followed) | ||