| Line 1: | Line 1: | ||
| + | ======Lightning Shutdown, Short Version: Updated 2/ | ||
| + | |||
| + | ===Changes coming out of Downtime 2018=== | ||
| + | - We have an " | ||
| + | - BokCCD2 is now the main Azcam computer for both the B&C and 90Prime instruments. | ||
| + | - There is no longer an UPS under the counter feeding console power. | ||
| + | - The weather computer is now password protected, and the password is posted on the monitor. | ||
| + | |||
| + | |||
| ===== CASS (B&C, WIFIS, SCC) Lightning Shutdown ===== | ===== CASS (B&C, WIFIS, SCC) Lightning Shutdown ===== | ||
| Line 12: | Line 21: | ||
| - Turn off IIS readout and unplug | - Turn off IIS readout and unplug | ||
| - Turn off IIS motor controller and unplug | - Turn off IIS motor controller and unplug | ||
| - | | + | - Turn off and unplug ' |
| + | | ||
| - Make sure everything is turned off and unplugged from orange outlets on the bottom of the telescope | - Make sure everything is turned off and unplugged from orange outlets on the bottom of the telescope | ||
| - Make sure IIS readout and motor controller are turned off and unplugged. | - Make sure IIS readout and motor controller are turned off and unplugged. | ||
| Line 28: | Line 38: | ||
| * Unplug dome mic if in use next to AED | * Unplug dome mic if in use next to AED | ||
| - Observers Workstation | - Observers Workstation | ||
| - | * Shutdown AzCam program, shutdown Azcam computer | + | * Shutdown AzCam program, shutdown Azcam computers |
| * Shutdown programs on Bart, shutdown Bart from Start Menu, and Bart Xterm | * Shutdown programs on Bart, shutdown Bart from Start Menu, and Bart Xterm | ||
| * Shutdown weather computer | * Shutdown weather computer | ||
| -Electronics Computer Rack Area | -Electronics Computer Rack Area | ||
| - | * Using the KVM switch make sure all computers in the rack are off: | + | * Using the KVM switch make sure all computers in the rack are off (**except BOKPOP**): |
| - | * #1 Bokccd1 B& | + | * #2 Bokccd2 Azcam Login (Should already be off) |
| - | * #5 Bart (should already be off) | + | * #5 Priscilla |
| - | * #6 Priscilla | + | * #6 Bart (should already be off) |
| - | * #7 BokPOP | + | * **#7 BokPOP |
| - | * Shut down Bokaux computer | + | |
| * Turn off flat field lamp controller if in use | * Turn off flat field lamp controller if in use | ||
| * Turn off printer | * Turn off printer | ||
| - | * Turn off (2) UPS units in bottom of the rack | + | * Turn off (3) UPS units in bottom of the rack |
| - | * Note: The CONTROL ROOM UPS also powers the Operators Console, make sure everything in it is turned off first. | + | * Note: The CONSOLE |
| Line 47: | Line 56: | ||
| - Unplug three plugs with red tape on your right as you enter the door down low | - Unplug three plugs with red tape on your right as you enter the door down low | ||
| - Unplug 1 twist lock plug with red tape west wall | - Unplug 1 twist lock plug with red tape west wall | ||
| - | - Unplug | + | - Unplug |
| *Second Floor Pier Room | *Second Floor Pier Room | ||
| Line 55: | Line 64: | ||
| - Unplug power strip from wall | - Unplug power strip from wall | ||
| - Disconnect phone line from fax machine | - Disconnect phone line from fax machine | ||
| + | - There is a computer tower that sits to the left of the desk. If on, turn off by pushing the power button. | ||
| Line 80: | Line 90: | ||
| * Shutdown weather computer | * Shutdown weather computer | ||
| -Electronics Computer Rack Area | -Electronics Computer Rack Area | ||
| - | * Using the KVM switch make sure all computers in the rack are off | + | * Using the KVM switch make sure all computers in the rack are off **except BokPOP** |
| - | * #2 Bokccd2 | + | * #2 Bokccd2 |
| - | * #4 Bokccd4 | + | * #4 Bokccd4 |
| - | * #4 Bart (should already be off) | + | * #5 Priscilla (if in use) |
| - | * #7 BokPOP | + | * #6 Bart (should already be off) |
| - | | + | * #7 BokPOP **(should stay ON)** |
| * Turn off flat field lamp controller if in use | * Turn off flat field lamp controller if in use | ||
| * Turn off printer | * Turn off printer | ||
| - | * Turn off (2) UPS units in bottom of the rack | + | * Turn off (3) UPS units in bottom of the rack |
| - | * * Note: The CONTROL ROOM UPS also powers the Operators Console, make sure everything in it is turned off first. | + | * * Note: The CONSOLE |
| *Warm Room Cable Closet | *Warm Room Cable Closet | ||
| - Unplug three plugs with red tape on your right as you enter the door down low | - Unplug three plugs with red tape on your right as you enter the door down low | ||
| - Unplug 1 twist lock plug with red tape west wall | - Unplug 1 twist lock plug with red tape west wall | ||
| - | - Unplug GPS Ethernet cable, | + | - Plug in Brandywine NTP Server' |
| *Second Floor Pier Room | *Second Floor Pier Room | ||
| Line 102: | Line 112: | ||
| - Unplug power strip from wall | - Unplug power strip from wall | ||
| - Disconnect phone line from fax machine | - Disconnect phone line from fax machine | ||
| + | - There is a computer tower that sits to the left of the desk. If on, turn off by pushing the power button. | ||
| =====Cold Start up- Coming out of Lightning Shutdown===== | =====Cold Start up- Coming out of Lightning Shutdown===== | ||
| Line 107: | Line 118: | ||
| - Plug in power strip | - Plug in power strip | ||
| - Connect phone line to fax machine | - Connect phone line to fax machine | ||
| + | - Turn on the computer tower to the left of the desk by pressing the power button. | ||
| *Second Floor Pier Room | *Second Floor Pier Room | ||
| Line 114: | Line 126: | ||
| -Plug in three plugs with red tape on your right as you enter the door down low | -Plug in three plugs with red tape on your right as you enter the door down low | ||
| -Plug in 1 twist lock plug with red tape west wall | -Plug in 1 twist lock plug with red tape west wall | ||
| - | -Plug in GPS Ethernet cable, | + | -Plug in Brandywine NTP Server' |
| *Electronics Computer Rack Area | *Electronics Computer Rack Area | ||
| - | - Turn on (2) UPS units in bottom of the rack | + | - Turn on (3) UPS units in bottom of the rack. On the new Console UPS: You will need to hit the power button, then use the arrow and OK buttons to choose "On Immediately" |
| - | - Turn on Bokpop | + | |
| + | - Turn on Bokpop | ||
| - Turn on Bart | - Turn on Bart | ||
| - | - Turn on BokAux | + | - Turn on BokAux |
| - Turn on printer | - Turn on printer | ||
| - **90Prime users** | - **90Prime users** | ||
| - | | + | |
| - | | + | |
| - **B&C users** | - **B&C users** | ||
| - | | + | |
| *Observers Work Station | *Observers Work Station | ||
| - | - Turn on Azcam Xterm, log into BokCCD1 | + | - Turn on Azcam Xterm, log into BokCCD2 |
| - | - Start Bart Xterm, | + | - Start Bart Xterm, |
| + | - Start weather computer by pressing button on top of the upper left edge. Enter the password posted on the monitor. | ||
| * At the Operators Console: | * At the Operators Console: | ||
| - | | + | |
| - Pull out emergency stop button on console | - Pull out emergency stop button on console | ||
| - Turn on TCS computer | - Turn on TCS computer | ||
| Line 153: | Line 167: | ||
| - Plug in IIS readout and turn on | - Plug in IIS readout and turn on | ||
| - Plug in IIS motor controller and turn on | - Plug in IIS motor controller and turn on | ||
| + | - Plug in and turn ON " | ||
| | | ||
| | | ||