-=====Lightning Shutdown-Long Version=====+=====Lightning Shutdown-Long Version, Updated 2/15/19=====
 This version is ideal for someone who is not as familiar with the lightning shutdown procedure.  This has ALL the steps and locations of computers.  For someone who is very familiar with lightning shutdown at the 90" and who observes frequently, you might like the short version which is linked on the previous page. This version is ideal for someone who is not as familiar with the lightning shutdown procedure.  This has ALL the steps and locations of computers.  For someone who is very familiar with lightning shutdown at the 90" and who observes frequently, you might like the short version which is linked on the previous page.
 +
 +==Changes Coming Out of Downtime 2018===
 +
 +  - We have an “always on” UPS now, which feeds a power strip above BokAux. The Aux computer, the WIFI, BokPOP, and the phones are all plugged into this power strip which means these are now always left on. You can use BokAux now to keep an eye on the weather by pulling up the weather sites bookmarked on the computer in Firefox.
 +  - BokCCD2 is now the main Azcam computer for both the B&C and 90Prime instruments. We are going to have a spare Azcam Login machine (BokCCD4) in the future, but it is now currently at ITL. The username and password to get in are the same as before.
 +  - There is no longer an UPS under the counter feeding console power. There is a new UPS in the black computer rack labeled “Console UPS” To shut this down you press the on/off button and select “Off Immediately” with the OK button. This takes a while to shut off.  When you turn it on you must select "On Immediately" with the OK button and arrows, otherwise the console will not turn on.  
 +  - The weather computer is now password protected, and the password is posted on the monitor. You turn on the weather computer by pressing the top edge of the left hand corner (it's beveled at a 45 degree angle), of the PC mounted on the wall next to the monitor.
 +
 +
 +
 ====Control Room: Telescope Control Workstation====  ====Control Room: Telescope Control Workstation====
 {{:a.jpg?nolink&300|}} {{:a.jpg?nolink&300|}}
Line 48: Line 58:
       -Click on "Primefocus Observer" or "Bok Observer" in upper right hand corner of the screen.  Click "Shutdown" in drop down menu.  This will shut down Bart.         -Click on "Primefocus Observer" or "Bok Observer" in upper right hand corner of the screen.  Click "Shutdown" in drop down menu.  This will shut down Bart.  
       -Click on the HP icon and select “PowerOff”       -Click on the HP icon and select “PowerOff”
 +    -Turn off weather computer cleanly using computer's start menu.
 ====Control Room: Near BokAux & KVM computer==== ====Control Room: Near BokAux & KVM computer====
 {{:o.jpg?nolink&300|}} {{:o.jpg?nolink&300|}}
Line 54: Line 65:
     NOTE: The 90" network now has a section that remains on during a lightning shutdown. This Always-ON-Network includes an UPS behind the wooden wall and several devices including BOKPOP and the AUX-PC.  Please leave these devices powered on when doing a lightning shutdown.     NOTE: The 90" network now has a section that remains on during a lightning shutdown. This Always-ON-Network includes an UPS behind the wooden wall and several devices including BOKPOP and the AUX-PC.  Please leave these devices powered on when doing a lightning shutdown.
 \\ {{:p.jpg?nolink&300|}}   \\ {{:p.jpg?nolink&300|}}  
-    -The number associated with “Bokccd2 90Prime & BCSpec” computer is “2”  +    -The number associated with “Bokccd2 Azcam Login” computer is “2”  
-    -The number associated with “Bokccd4 90Prime & BCSpec spare” is “4” +    -The number associated with “Bokccd4 Spare Azcam Login” is “4” 
     -The number associated with "Pricilla" is "5"-this most likely will not be on.     -The number associated with "Pricilla" is "5"-this most likely will not be on.
     -The number associated with “Bart” is “6”      -The number associated with “Bart” is “6” 
     -The number associated with "bokpop" is "7" (Please leave BOKPOP turned ON)     -The number associated with "bokpop" is "7" (Please leave BOKPOP turned ON)
   -Turn off “Ring Lamps” Flat Field Lamp Controller.    -Turn off “Ring Lamps” Flat Field Lamp Controller. 
-  -Turn off “Platform Lamps” Flat Field Lamp Controller.\\ {{:q.jpg?nolink&300|}}\\    +  -Turn off “Platform Lamps” Flat Field Lamp Controller.\\ {{:q.jpg?nolink&300|}}\\    
-  -Close any open programs or windows & shutdown the “bokaux” computer. +
   -Turn off the laser printer.    -Turn off the laser printer. 
   -At the bottom of the electronics/computer rack, turn off the following THREE (3) UPS units right below the lamp controller:   -At the bottom of the electronics/computer rack, turn off the following THREE (3) UPS units right below the lamp controller:
     -INSTRUMENT UPS  (press the recessed "O" button to turn OFF)     -INSTRUMENT UPS  (press the recessed "O" button to turn OFF)
     -CONTROL ROOM UPS  (press the recessed "O" button to turn OFF)     -CONTROL ROOM UPS  (press the recessed "O" button to turn OFF)
-    -CONSOLE UPS  (press the red button and then the OK button to select shut off immediately)   +    -CONSOLE UPS  (press the power button and then the OK button to select shut off immediately)   
 ====Control Room: Closet Behind Computer Rack==== ====Control Room: Closet Behind Computer Rack====
 Access to this area is through the side door (labeled “Warm Room Cable Closet”) located between the Bok Walk and south warm room door. All cables and AC cords to be disconnected are identified by red tape. Access to this area is through the side door (labeled “Warm Room Cable Closet”) located between the Bok Walk and south warm room door. All cables and AC cords to be disconnected are identified by red tape.
Line 92: Line 102:
     - Turn on computer to the left of the desk on the floor.     - Turn on computer to the left of the desk on the floor.
   - Second Floor Pier Room   - Second Floor Pier Room
-    - Plug in UPS and turn on+    - Plug in UPS and turn on by pressing (I) button
   - Warm Room Cable Closet   - Warm Room Cable Closet
     -Plug in three plugs with red tape on your right as you enter the door down low     -Plug in three plugs with red tape on your right as you enter the door down low
Line 98: Line 108:
     -Plug in Brandywine NTP Server's Ethernet cable, BNC cable and power cable on west wall     -Plug in Brandywine NTP Server's Ethernet cable, BNC cable and power cable on west wall
   - Electronics Computer Rack Area   - Electronics Computer Rack Area
-    - Turn on (3) UPS units in bottom of the rack+    - Turn on (3) UPS units in bottom of the rack. New Console UPS-press the power button then select "On Immediately" with the arrows and OK button or it will not turn on. Power button will go from red to white.
     - Turn on Bokpop (it should be already ON)     - Turn on Bokpop (it should be already ON)
     - Turn on Bart     - Turn on Bart
-    - Turn on BokAux+    - Turn on BokAux (it should already be on)
     - Turn on printer     - Turn on printer
     - **90Prime users**     - **90Prime users**
Line 110: Line 120:
   - Observers' Work Station   - Observers' Work Station
      - Turn on Azcam Xterm, log into BokCCD2 if using B&C or 90Prime.  Start the Azcam program.        - Turn on Azcam Xterm, log into BokCCD2 if using B&C or 90Prime.  Start the Azcam program.  
-     - Start Bart Xterm, log in either as Primefocus Observer if you're using 90Prime, or Bok Observer with any other instrument using appropriate passwords.  Start programs on Bart.  +     - Start Bart Xterm, log in either as Primefocus Observer if you're using 90Prime, or Bok Observer with any other instrument using appropriate passwords.  Start programs on Bart.  
 +     - Turn on weather computer by pressing the button on top of the upper left corner. 
   - At the Operators Console:   - At the Operators Console:
      - Make sure the CONTROL ROOM UPS in the Electronics Computer Rack is plugged in and turned ON (should be done already if the above steps were followed)      - Make sure the CONTROL ROOM UPS in the Electronics Computer Rack is plugged in and turned ON (should be done already if the above steps were followed)