=====Lightning Shutdown===== | =====Lightning Shutdown===== | ||
- | - Complete "End of Normal Night of Observing" | + | - Complete "End of Normal Night of Observing" |
- In Dome | - In Dome | ||
- | * Turn off all instrument electronics (for M4K this would be camera power supply, guider controller, filter wheel, guider chiller, shutter control box). | + | * Turn off all instrument electronics (for M4K this would be the camera power supply, guider controller, |
- | * Unplug the four (4) cords from the “clean” orange outlets on the underside of the telescope. The cords go to the camera power supply and guider controller | + | * Unplug the four (4) cords from the “clean” orange outlets on the underside of the telescope. The cords from the eastside outlets |
* Shut down laptop computer (if in use) and unplug two (2) cords from the “clean” orange UPS outlets located near the floor on the platform control console. | * Shut down laptop computer (if in use) and unplug two (2) cords from the “clean” orange UPS outlets located near the floor on the platform control console. | ||
- | * Turn off TCS by turning the rotary switch on the side of the junction box. NOTE: NEVER TURN OFF TCS UNLESS TELESCOPE IS IN STOW POSITION! | + | * Turn off TCS by turning the red rotary switch on the side of the TCS junction box. NOTE: NEVER TURN OFF TCS UNLESS TELESCOPE IS IN STOW POSITION! |
* Lower platform (if not already lowered). | * Lower platform (if not already lowered). | ||
* Turn off telescope power and main power (note instructions on console regarding order of switches). | * Turn off telescope power and main power (note instructions on console regarding order of switches). | ||
Line 13: | Line 13: | ||
* Shut down BigPop computer and Telcom Server via button or switch on front of their panels. | * Shut down BigPop computer and Telcom Server via button or switch on front of their panels. | ||
* Turn off printer. | * Turn off printer. | ||
- | * Unplug all power cords marked with red tape from “clean” (orange) outlets on wall behind the computer monitors. | + | * Unplug all power cords marked with red tape from orange outlets on wall behind the computer monitors. |
* Turn off CSS surge strip at back of table by the door and unplug surge strip cord from orange outlet below (if not already done). | * Turn off CSS surge strip at back of table by the door and unplug surge strip cord from orange outlet below (if not already done). | ||
- In UPS room | - In UPS room | ||
Line 19: | Line 19: | ||
* Unplug GPS (Brandywine) and yellow power strip cords from “clean” (orange) power outlet (both are marked with red tape). | * Unplug GPS (Brandywine) and yellow power strip cords from “clean” (orange) power outlet (both are marked with red tape). | ||
* Unplug Trendnet (small white box w/red tape and grey network cable attached) from “dirty” (regular) power outlet (NOTE: Trendnet MUST be plugged into a “dirty” power outlet when coming out of lightning shutdown). | * Unplug Trendnet (small white box w/red tape and grey network cable attached) from “dirty” (regular) power outlet (NOTE: Trendnet MUST be plugged into a “dirty” power outlet when coming out of lightning shutdown). | ||
- | * Unplug any cords plugged into the orange outlets above the UPS, in particular cord (marked with red tape) for the CSS rack | + | * Unplug any cords plugged into the orange outlets above the UPS, in particular |
* Follow separate instructions for UPS shutdown posted on UPS room wall. | * Follow separate instructions for UPS shutdown posted on UPS room wall. |