 =====Lightning Shutdown===== =====Lightning Shutdown=====
  
-  - Stow the telescope and dome +  - Complete "End of Normal Night of Observing" shutdown
-    * TCS-NG-INDI -> Goto Functions -> Stow +
-    * DOME-NG-INDI -> Dome Actions -> Stow +
-  - Disable the telescope: TCS-NG-INDI -> Actions -> Enable +
-    * Note: green light will turn red +
-  - Close the Mirror Cover +
-  - Close the Dome slit +
-  - Close the Wind Screens+
   - In Dome   - In Dome
-    * Turn off all instrument electronics, including the camera shutter, and unplug from orange outlets under scope (Four available outlets)+    * Turn off all instrument electronics, including the camera shutter, and unplug two power strips and CCD power cord from orange plugs under the telescope.  
 +    * On the west side, two power strips are now plugged into the two outlets.  Those are what need to be unplugged for lightning shutdown on this side  
 +    * If in use: shut down laptop computer, and unplug laptop from the orange UPS outlets located near the floor on the platform control console 
 +    * The focus is no longer operating off the orange extension cord that plugged into the powerstrip on the south end of the scope, it’s hardwired in, so you don’t have to worry about it. 
     * Turn off TCS computer using the light switch     * Turn off TCS computer using the light switch
-    * Turn off Telescope power +    * Turn off telescope power 
-    * Lower Platform+    * Lower platform
     * Turn off main power     * Turn off main power
-    * Turn off and unplug laptop in dome 
   - In Control room   - In Control room
-    * On BIGAUX double click "shutdown bigpop", and click Shutdown bigpop when prompted+    * On BIGAUX double click "shutdown bigpop", and click shutdown bigpop when prompted.  You will need to enter the username "bigobs" and the Gerard password.  
     * Manually shutdown all computers.      * Manually shutdown all computers. 
       * Gerard       * Gerard
Line 23: Line 18:
       * Bigguider       * Bigguider
       * BIGAUX       * BIGAUX
-    * Turn Off printer.+    * Turn off printer. 
 +    * Turn off Telcom server
     * Turn off and unplug power strip on west wall.     * Turn off and unplug power strip on west wall.
     * Unplug printer relay.     * Unplug printer relay.
-    * Unplug everything from orange outlets behind observer work station +    * Unplug everything from orange outlets behind observer work station, as well as UPS outlets by the video finder scope monitor
-    * Turn off Telcom server box.+
     * Unplug computer racks (plugged into orange outlet #4 on wall below emergency stop)     * Unplug computer racks (plugged into orange outlet #4 on wall below emergency stop)
-    * Unplug video finderscope from UPS power under the old stereo system. +  - In UPS Room, unplug
-  - In UPS Room: +    The emergency stop plug (bright yellow) no longer needs to be unplugged when going into lightning shutdown. 
-      Unplug GPS +    GPS 
-      Unplug Emergency Stop Power 110 VAC +    * Shutdown UPS following instructions posted on board.
-      *Shutdown UPS following instructions posted on board.+